Pathology Service Reading Out Biopsy Slides in Academia

Abstract

Remote digital pathology allows healthcare systems to maintain pathology operations during public health emergencies. Existing Clinical Laboratory Improvement Amendments regulations require pathologists to electronically verify patient reports from a certified facility. During the 2019 pandemic of COVID-19 disease, acquired past the SAR-CoV-two virus, this requirement potentially exposes pathologists, their colleagues, and household members to the risk of becoming infected. Relaxation of government enforcement of this regulation allows pathologists to review and written report pathology specimens from a remote, non-CLIA certified facility. The availability of digital pathology systems can facilitate remote microscopic diagnosis, although formal comprehensive (example-based) validation of remote digital diagnosis has not been reported. All glass slides representing routine clinical signout workload in surgical pathology subspecialties at Memorial Sloan Kettering Cancer Center were scanned on an Aperio GT450 at ×40 equivalent resolution (0.26 µm/pixel). Twelve pathologists from ix surgical pathology subspecialties remotely reviewed and reported consummate pathology cases using a digital pathology organisation from a not-CLIA certified facility through a secure connection. Whole slide images were integrated to and launched within the laboratory information system to a custom vendor-agnostic, whole slide paradigm viewer. Remote signouts utilized consumer-grade computers and monitors (monitor size, 13.iii–42 in.; resolution, 1280 × 800–3840 × 2160 pixels) connecting to an institution clinical workstation via secure virtual private network. Pathologists afterwards reviewed all respective drinking glass slides using a light microscope within the CLIA-certified department. Intraobserver cyclopedia metrics included reporting elements of top-line diagnosis, margin status, lymphovascular and/or perineural invasion, pathology stage, and coincident testing. The median whole slide prototype file size was one.3 GB; scan time/slide averaged 90 south; and scanned tissue area averaged 612 mm2. Signout sessions included a full of 108 cases, comprised of 254 individual parts and 1196 slides. Major diagnostic equivalency was 100% between digital and glass slide diagnoses; and overall concordance was 98.8% (251/254). This study reports validation of master diagnostic review and reporting of complete pathology cases from a remote site during a public health emergency. Our feel shows high (100%) intraobserver digital to glass slide major diagnostic concordance when reporting from a remote site. This randomized, prospective written report successfully validated remote utilise of a digital pathology organisation including operational feasibility supporting remote review and reporting of pathology specimens, and evaluation of remote access performance and usability for remote signout.

Introduction

The digital transformation of pathology has allowed for digitization of glass slides to generate whole slide images (i.eastward., WSIs or digital slides). The primary components of a digital pathology system include a whole slide scanner, software whole slide image viewer, and display monitor. Digital pathology allows pathologists to access, evaluate, and share pathology slides using a digital workflow, and provides facile access for remote sign-out, when advisable. It offers pathologists a novel method to review and render diagnoses, broadens clinical expertize, and allows for digital direction of pathology slides for reporting of patient specimens. An affluence of literature has been published to show cyclopedia of rendering diagnoses using WSIs compared to glass slides [ane,2,3,four,5,6,7,8,9,ten,11,12,13,14,15,16,17,eighteen,xix,twenty,21,22,23,24,25]. These studies take shown that reporting of pathology specimens is readily achieved using WSIs and can exist incorporated into routine anatomic pathology laboratory workflow. Furthermore, digital pathology can exist leveraged to create and raise a digital workflow for pathologists. At cancer centers such every bit Memorial Sloan Kettering Cancer Heart (MSK), patients benefit from experienced sub-specialized expert pathologists. Digital pathology has also enhanced the pathologist's experience and has provided efficiency and operational savings [3].

Clinical Laboratory Comeback Amendments (CLIA) regulations require pathologists to electronically verify patient reports from the CLIA-certified facility [26]. Yet, during the COVID-19 pandemic, this requirement potentially exposed pathologists and trainees to the SARS-CoV-2 virus and thus put the pathologist, colleagues, and household members at hazard of infection from a potentially lethal virus. Moreover, in any public wellness emergency, if pathologists are unable to be present at the CLIA-licensed facility due to affliction, quarantine, or other travel restrictions, hospital systems will lose continuous patient care and delay turnaround times of pathology reporting. Many pathologists have secure access (e.g., virtual private networks) through hospital firewalls and tin connect to hospital information systems (i.eastward., laboratory data systems, electronic medical records) from remote locations (e.chiliad., dwelling house). During this time, with the advent of digital pathology, novel digital workflows tin can be implemented and validated for remote utilise.

This written report presents a validation of the digital pathology system at a big academic center in New York Metropolis, the world-wide epicenter of the COVID-19 pandemic. The validation encompasses digitization of glass slides generated from formalin fixed alkane embedded and frozen tissue; and includes hematoxylin & eosin stains, immunohistochemical stains, and special stains. Glass slides included those generated in the MSK histology laboratory equally well as slides received every bit consultation cases from referring institutions. The whole slide imaging process includes pre-analytic quality assurance of the glass slides, the analytic process of placing the glass slides on the whole slide scanner for scanning, and post-analytic quality balls of the generated whole slide images. The whole slide scanning process includes the scanner capturing images from a glass slide nether high resolution and stitching those images together to form a digital file that can be navigated similarly to a glass slide on a microscope, while using a digital workflow. WSI software viewers are developed to view the WSIs for pathologic diagnosis. At MSK, whole slide scanners from multiple vendors and custom viewer software haven been implemented and validated to support clinical workflow. The aims of this study include: (i) to appraise operational feasibility supporting remote review and principal reporting of pathology specimens using digital pathology systems; (2) to evaluate remote access functioning and usability for pathologist remote signout; (three) to validate a digital pathology organisation for remote primary diagnostic utilize.

Fabric and methods

Assessment of remote readiness

Prior to engaging a large-scale validation study, we sought to appraise the departmental readiness for the technical accessibility needed for remote signout. A survey was distributed to all clinical members in the department, including all faculty and trainees.

Questions in the remote readiness survey included:

  1. 1.

    Exercise y'all have a computer/workstation that you lot can employ to work from habitation?

  2. 2.

    What is the largest monitor on the reckoner that you tin can work on?

  3. iii.

    Do you have cyberspace at home, if and then, what is the network bandwidth?

  4. 4.

    Do you have institutional virtual private network and 2-cistron hallmark configured?

  5. 5.

    Do yous know how to connect your device to a hospital clinical workstation through remote desktop connection?

Case selection

Prospective selection of patient specimens included cases from our big bookish tertiary comprehensive Cancer Heart. Pathologists' clinical service days were randomly selected for remote digital signout. Each signout day was composed of all specimen classes including biopsies, surgical resections, and departmental consultation material (e.g., pathology material from other institutions) accessioned for a given signout per routine departmental protocol. Each pathologist received their corresponding case worklist on their randomly assigned clinical service day, comprised of that mean solar day's unabridged clinical volume within their subspecialty. Pathologists reviewed an entire day's workload including all slides and levels taken from biopsies, all immunohistochemical stains, and special stains submitted with consultation cases. Cases where frozen sections had been performed were reviewed together with the corresponding frozen section control slides equally per the pathology department routine workflow. A referee pathologist not participating in case signout verified all cases included in the report by ensuring all cases accessioned for each pathologist's signout session were scanned and assigned to pathologists' worklists.

Whole slide scanning protocol

Specimens received from within our hospital (hereafter referred to as "in-business firm") underwent conventional procedures for specimen accessioning, gross pathology evaluation, and autopsy of tissue into cassettes/blocks, tissue processing, embedding, microtomy, staining, and coverslipping. This study included all cases accessioned and set up for clinical review on each pathologist'due south randomly assigned clinical service 24-hour interval. The overall operational workflow from accessioning of the patient case through slide distribution to pathologists is shown in Fig. 1. Glass slide hematoxylin & eosin staining, coverslipping, and drying of mounting media were prepared by Leica Spectra instruments (Leica Biosystems, Buffalo Grove, Illinois, USA). The Spectra instruments use Leica Universal Slide racks, which can be directly inserted into an Aperio GT450 whole slide scanner (Leica Biosystems, Buffalo Grove, Illinois, USA). After glass slide staining and coverslipping, the glass slides are loaded into the Aperio GT450 by laboratory personnel. Glass slides were scanned at ×40 equivalent magnification (0.26 µm/pixel) using a native ×xl objective lens (Leica Biosystems).

Fig. ane: Operational workflow for digital pathology accessioning, staining, scanning, and slide distribution.
figure 1

The modified operational workflow illustrates the process and roles involved from the point of accessioning to slide distribution.

Full size image

For patient specimens received equally consultations from referring institutions, the drinking glass slides were generated at the referring establishment, shipped to MSK, and and so accessioned into the laboratory information system. A departmental consult barcode was affixed to the consult glass slide, which was subsequently placed in Leica Universal Slide racks at the bespeak of accessioning. Glass slide choice upwards and drop off stations were placed in the pathology department accessioning area and histology laboratory. Digital scanning team members routinely circulated to the accessioning area to courier racks of drinking glass slide consults to undergo quality control (see "Epitome Quality Control") and be placed on the Aperio GT450 whole slide scanner. All pathology avails (e.g., specimen container, blocks, slides) had 2D data-matrix barcodes. The barcode on the glass slide label enabled the Leica Aperio eSlide manager database to interface with the laboratory information arrangement (Cerner CoPathPlus). Whole slide images were accessed through the PICSPlus module in CoPathPlus and launched into a custom whole slide epitome viewer. Glass slides from accessioning and the histology laboratory were scanned, including, only not express to the subspecialties designated for the pathologists' signouts. The boosted drinking glass slides not used as part of the validation study were scanned and reviewed for quality assurance; and to perform a technical cess on the Aperio GT450 whole slide scanner.

Image quality command

All whole slide images were reviewed by technicians or the referee pathologist equally role of the quality assurance process prior to initiation of primary digital signout.

There was no change to the laboratory's electric current standard operating and quality command procedures for initial handling (accessioning), gross test or histologic preparation (processing, embedding, microtomy and staining) of surgical pathology specimens. The quality control procedure also involves macro-evaluation of pre-analytic slide artifacts to be resolved earlier scanning. The following guidelines were provided for quality control: glass slides should be appropriately stained and stale, non broken, and clean of ink markings; coverslips were to exist properly placed without hanging over the edge of the drinking glass slide, and air bubbles were to be absent-minded; the glass slide label should be apartment and non extending past the slide edge or covering tissue on the slides. For consult slides, this process included a barcoding and tracking station to scan each glass slide; this ensured that the boosted MSK laboratory barcode was readable by handheld barcode readers and marked the glass slide every bit scanned (received for scanning in the laboratory information organization). Consult slides without barcodes, or barcodes that failed reading by the handheld reader had new patient labels with barcodes printed and placed on the glass slide label area. Boosted existent-time quality control was performed by the whole slide scanner that provided user interface letters including barcode detection failures, no tissue detection, macro focus image failure, or image quality errors. Prior to unloading the glass slides, following glass slides scanning, some other quality control evaluation was performed by the digital imaging technologists to ensure the WSI was of adequate epitome quality. Prototype thumbnails were reviewed to ensure all tissue present on the drinking glass slide was scanned and wholly represented on the WSI, and confirmation the slide barcode was decoded and present within the LIS. The whole slide scanner touchscreen display interface provided firsthand feedback to the scanning operator who addressed whatever scanning errors immediately (e.g., rescan of a slide with an mistake). An additional mail service-scan quality control mensurate included the pathologist review of the WSI. The whole slide image viewer had a custom in-built tool to report a trouble with the digital slide, which sends a notification to the digital scanning team to rescan that glass slide (Fig. 2). The pathologist had discretion to enquire for a rescan of a glass slide if desired, or if whatsoever WSI revealed scanning artifacts.

Fig. 2: Pathologist whole slide paradigm viewer feedback and communication tool.
figure 2

Users have a drop down list (shown above) available to study whole slide epitome quality problems. Once selected, a notification is delivered to the Digital Scanning Team to rescan the intended slide.

Total size image

WSI viewer

The WSI viewer software is an institutional standard, vendor-agnostic, whole slide prototype viewer. As per the institution scanning protocol, prospective clinical cases have been digitally scanned and interfaced with the laboratory information arrangement. For over 3 years, all pathologists at our institution have used the WSI viewer on a daily footing and reviewed retrospectively archived whole slide images from pathology specimens. The WSI viewer launches from the PICSPlus module in Cerner CoPathPlus and visualizes any scanned whole slide images for that case (east.m., all, multiple, or individual); review of the WSI is controlled through the computer mouse and keyboard. The web-based viewer is initiated within the pathologist's default web browser. For pathologist navigation, the WSI viewer software includes tools such every bit thumbnail viewing, slide label viewing, zooming, panning, and co-registration of multiple slides. Annotation tools include a digital ruler for measurements, tracking of viewed slides, heatmap coverage of WSI regions reviewed, screenshots, and the power to add annotations.

Digital and glass signout

After each glass slide browse was successfully completed on the whole slide scanner, WSI are immediately available in the laboratory information system, automatically collated by instance and sequential slide order equally they exist in the laboratory data system. At our institution, pathology trainees (i.e., fellows) preview and compose a preliminary study for each patient case prior to reviewing with a kinesthesia pathologist for final reporting. For this study, trainees were included in the overall reporting workflow with interest in handling and previewing of the cases prior to pathologist's review. Pathology trainees received the glass slides with respective newspaper surgical pathology requisitions for each instance, as distributed from the primal laboratory. In addition, trainees were likewise provided with admission to the digital slides at the time of previewing and had the selection to utilize either glass or digital slides when previewing cases to facilitate social distancing efforts within the department. After trainees entered preliminary reports in the laboratory information system, cases were subsequently assigned to a pathologist for review. Pathologists independently reviewed all cases as digital slides first, from a remote site (e.g., home) via secure access through a virtual individual network and 2-factor authentication per institutional policy. Pathologists accessed clinical workstations located inside the institutional enterprise firewall and had access to hospital data systems such every bit the laboratory information system, electronic medical record, and radiology picture archiving and communication system from their remote site. Each clinical workstation was an institutional standard configuration (3.ii GHz, 8 GB random-access retention (RAM), 64-bit processor) with default spider web browser as Google Chrome or Microsoft Cyberspace Explorer 11. From the remote site, pathologists used the laboratory information organisation to view all surgical pathology requisitions digitally and launch WSIs into the WSI viewer. After all digital slides were reviewed for each case, a consummate concluding report was entered into the laboratory data organization, but not electronically released to the medical record from the remote site. Subsequent case re-review was performed in pathologists' offices (at the CLIA-licensed facility) using glass slides on an Olympus BX series calorie-free microscope (i.e., Olympus BX43). Pathologists re-reviewed all glass slides for all cases, mirroring the remote digital WSI review. The interval between initial remote digital review and secondary glass slide review was brusque (mean, 2 days) to conform with turnaround time reporting requirements. Pathologists reported all cases according to their routine workflow (i.e., complimentary-text, worksheet templates, synoptic worksheets) including all required diagnostic reporting elements. Prior patient pathology reports and slides (from prior patient'due south pathology) were made available to the pathologists for both signout sessions via previously archived WSIs. Concordance metrics were captured after each digital and glass slide signout. Other data collected included remote hardware utilized, such as: network bandwidth while connected to virtual private network, web-browser (for spider web-based WSI viewer), monitor size, brandish resolution, estimator processing unit, and RAM. Technical difficulties were besides recorded, if whatever. Later both signout sessions, an experience survey was distributed to the pathologists who participated in remote digital and on-site glass slide review.

Questions in the reader experience survey included:

  1. 1.

    How many years have you been practicing pathology?

  2. 2.

    How many years of experience do you accept using digital pathology? (in any capacity)

  3. 3.

    Charge per unit the digital pathology slide viewer.

  4. 4.

    Rate your satisfaction with the launching of slides from within the laboratory information organisation (CoPath).

  5. 5.

    Rate the quality of the digital slides.

  6. six.

    Rate your satisfaction with the performance in navigating the digital slides.

  7. 7.

    How comfy would you experience providing primary diagnosis using digital pathology, with retrieval of glass slides available upon request?

  8. 8.

    How comfy would you feel providing primary diagnosis using digital pathology, without availability of glass slides?

Concordance

A read is defined in this study equally a diagnosis for a specimen part (i.due east., for each case, there may be i or multiple parts, each comprising one or multiple slides). Data captured after each signout session included, subspecialty, specimen type (i.e., biopsy, resection, in-firm, consultation), part blazon description (i.eastward., organ site), number of slides per case, and ordered ancillary tests (eastward.g., hematoxylin & eosin stained recuts, immunohistochemical and special stains). Recorded diagnostic data included top-line diagnosis, margin status, lymphovascular and perineural invasion, pT and pN stage; and any changes in estimation between remote digital and on-site glass slide signout sessions. Cyclopedia metrics were evaluated based on paired reads, 1 read by WSI and one read by brightfield microscopy for the same case by the same pathologist. Frequency of concordant/discordant diagnoses were calculated. The drinking glass slide diagnosis that was ultimately reported was considered the reference (i.eastward., golden standard diagnosis). The laboratory's defined concordance thresholds were based on criteria used by other well-established studies examining WSI primary diagnosis using Food and Drug Administration cleared digital pathology systems [2, 3, 27, 28]:

  • Diagnoses rendered past WSI & drinking glass slides accept a major discrepancy rate of 4% or less

  • The major discrepancy rate of the diagnoses rendered on WSI (relative to the reference diagnosis) was 7% or less.

Major diagnostic discrepancies were divers as any finding identified by ane modality and not identified on the other that would exist clinically significant. Minor diagnostic discrepancies were defined equally any finding identified past one modality and not identified in the other that would not touch on clinical care. A pathologist who did not participate as a reader in this study adjudicated concordance of diagnoses.

Results

Department readiness for remote signout

The readiness survey was distributed to all clinical kinesthesia and trainees (n = 142); there were 112 respondents.

Of the respondents, 96% (108/112) had a figurer to perform remote work. The distribution of monitor size available on the reckoner was <14 in. (41.seven%, 43/103); 15–20 in. (31.1%, 32/103); 21–25 in. (17.5%, 18/103); and >25 in. (9.7%, 10/103). 92.5% (99/107) of respondents had access to net at their remote site. Median network bandwidth (e.m., download speed) was 94 megabits per second (range 3–835). Network download speeds were aggregated as follows: <20 (13%, 10/76); 20–49 (11%, 8/76); fifty–99 (32%, 24/76); ≥100 (45% 34/76) megabits per second. Most (88%, 84/96) respondents had virtual private network connections configured with 2-gene hallmark setup, and 77% (63/82) knew how to appropriately connect the remote device to their hospital clinical workstation through remote desktop connexion. All faculty pathologists included in this validation report had access to a remote workstation with virtual individual network, 2-gene hallmark, and remote desktop connection.

Pre-analytic data

For the purposes of whole slide scanner technical evaluation, a full of 2119 glass slides were included and scanned. Of the 2119 WSI, the median scan area was 612 mm2. The median scan fourth dimension per slide at ×40 equivalent resolution (0.26 µm/pixel) was 90 due south. The median file size was one.three gigabytes (GB).

Quality balls review of WSIs past technicians or the referee pathologist prior to initiation of primary digital signout revealed that, of the 2119 drinking glass slides, 39 required rescanning due to barcode detection failures (n = 21), no tissue detected (n = 1), tissue detection failure (n = 9), no macro image focus (n = 1), and epitome quality (n = 7). The overall rescan charge per unit was 1.84%. These rescanned slides were farther analyzed past differentiating glass slides prepared in the MSK histology laboratory (northward = 13/1573; 0.83%) compared to those glass slides received in consultation (n = 26/546; 4.76%) (Table 1). Standard procedure for consult labeling includes affixing a secondary MSK Section consult label directly below the referring institution characterization. At the offset, the bulk of barcode detection failures were from consult glass slides during the scanning for the outset two pathologists' remote signout sessions. These errors were addressed past the digital pathology organisation vendor, who successfully deployed a feature inside the organisation to decode only MSK laboratory-specific barcodes. Consult drinking glass slide barcode detection failures after implementation of the software update decreased by 46% for all subsequent scanning of consult slides, 2.59% (n = 4/154). After initial image quality control of scanned slides, no WSIs were requested to be rescanned by the pathologists during the reader sessions.

Table 1 Technical evaluation of Aperio GT450 glass slide rescans.

Full size table

Twelve subspecialized surgical pathologists ("readers"), reported on cases from nine dissimilar surgical pathology subspecialties; with 1 reader for each and two for breast and genitourinary cases. A total of 16 reader sessions were conducted; one reader participated in three signout sessions, and two readers participated in two signout sessions (i.e., 16 remote and xvi on-site slide reader sessions). Surgical pathology subspecialties included: chest, bone & soft tissue, dermatopathology, gastrointestinal, genitourinary, gynecologic, head & neck, neuropathology, and thoracic pathology. Table 2 refers to the case, part, and slide distribution past reader. A full of 108 unique patient cases, 254 individual parts, and 1196 slides were included in the report. The readers completed 508 reads (i.e., diagnoses by part) in total: 254 reads past remote digital microscopy, and 254 reads using drinking glass slides on premises. Patient example distribution by specimen type included 204 in-house parts (i.e., biopsies and resections performed at our institution) and 50 parts sent for consultation from other institutions. All slides from each case were included for digital and drinking glass slide review. There was a hateful of 11 slides per case (range, 1–49). Specimen types in this patient cohort included a standard distribution of cases as typically expected in a large tertiary intendance center with a main focus on cancer (Table 3).

Table 2 Case, read, and slide distribution by pathologist (reader).

Full size table

Tabular array iii List of specimens in each respective subspecialty.

Total size table

Technical and hardware specifications

Pathologists' employ of remote workstations ranged from consumer grade laptop computers to college specification desktop computers with loftier definition dual monitors. Two pathologists reported virtual individual network errors (i.e., loss of connectivity), but were able to immediately reconnect without additional errors. Network bandwidth for readers in this written report while connected to the virtual private network ranged from 20 to 849 megabits per 2d. xi of the xvi pathologist remote sessions included a single monitor ranging from 13.3 to 25 in. and 1280 × 800 to 3000 × 2000 pixel resolution. The other five pathologist remote reader sessions had two monitors where the monitor size and brandish resolutions ranged from 13.5 to 42 inches and 1900 × 1080 to 3840 × 2160 pixel resolution. All monitors had 24-chip color depth. Calculator processing units ranged from 1.three–3.two GHz. No departure in WSI latency was identified by remote workstation but was reportedly slower with lower network connectivity. Of the 16 reader sessions, 69% (eleven/xvi) of readers used Google Chrome spider web browser, and 31% (v/xvi) used Internet Explorer eleven.

Concordance

A total of 508 reads were performed using both whole slide image and glass slides, 254 reads each. From all matched glass slide and whole slide image diagnoses, at that place was 100% major diagnostic cyclopedia (254/254 reads). Iii small-scale diagnostic discrepancies occurred, corresponding to an overall 98.8% (251/254 reads) for glass slide (at CLIA-certified facility) to remote WSI diagnostic concordance rate. Even readers with network speeds of less than 50 megabits per second had 100% major cyclopedia between digital and glass diagnostic reads. Diagnostic discrepancies were defined as any finding identified past one modality and not identified on the other that would be clinically meaning. The three discrepant diagnoses are detailed in Table iv. I each occurred in the breast, thoracic, and caput & neck subspecialties. Size measurements performed by digital annotation were 3% larger than compared with the glass slide measurements. Cyclopedia for presence and absence of lymphovascular and/or perineural invasion overall was 96.3% (26/27) for each appropriate part-level diagnosis. The surgical margin status, pathology tumor and node stage (east.g., pT, pN) were 100% concordant. Upon review of the discrepant cases, the diagnoses based on glass slide reads were deemed to be correct. For the iii small discrepant reads, after conclusion of the study, retrospective re-review of the whole slide images identified the pathologic findings non reported during remote digital review. Table five summarizes the cyclopedia evaluation.

Table 4 Discordant cases between whole slide image and glass slide reads.

Full size table

Table 5 Cyclopedia betwixt whole slide prototype and drinking glass slide reads for all reader sessions.

Full size tabular array

Reader experience survey

The reader feel survey distributed after completion of both digital and glass slide signout sessions produced ten responses (Fig. 3).

Fig. 3: Remote digital pathology experience survey.
figure 3

Responses to the digital pathology experience survey showed a broad range in years of pathology practice and years using digital pathology. Of note, two pathologists had more years of feel using digital pathology (i.e., during residency) than practicing every bit board certified pathologists. The majority of respondents rated their experience positively. Left: (i) How many years accept you been practicing pathology? (2) How many years of experience practice y'all have using digital pathology (in whatever capacity)? Right: (3) Charge per unit the digital pathology slide viewer. (four) Rate your satisfaction with the launching of slides from within the laboratory information system (CoPath). (v) Rate the quality of the digital slides. (6) Charge per unit your satisfaction with the performance in navigating the digital slides. (7) How comfortable would you feel providing primary diagnosis using digital pathology, with retrieval of glass slides available upon asking? (8) How comfy would you feel providing primary diagnosis using digital pathology, without availability of drinking glass slides?

Full size image

Of the ten survey respondents, the readers had practiced pathology for a median of 19 years (range iii–37). All readers had some feel using digital pathology, with the median number of years existence 5 years (range 3–x). On a 5-point Likert scale (one—very poor, 5—very good), all users rated the WSI Slide Viewer equal to or >3, with a median rating of four (i.e., good). The reader satisfaction with launching whole slide images from the laboratory information system was ≥4 by all readers. Regarding the whole slide images, readers reported the median prototype quality rating equally 5 (i.due east., very adept). The satisfaction with the performance in navigating the whole slide images varied amidst the readers: v responded as 3 (i.e., neutral), three respondents every bit 4 (i.e., good), and two readers every bit v (i.due east., very good). Comfort level regarding the option of signout using whole slide images for primary diagnosis with the availability of glass slides had a median rating of five, where 90% of readers responded ≥four (i.east., skilful), and simply one reader as three (i.e., neutral). When asked their level of condolement regarding principal diagnosis using digital pathology without the availability of drinking glass slides, the median reader rating was agreeable; 1 reader rated every bit 2 (i.e., uncomfortable), three readers equally 3 (i.e., neutral), 5 readers equally (i.e., comfortable), and one reader as 5 (i.e., very comfy). Comments from pathologists included demand for a better input device (e.g., computer mouse) for slide navigation. Readers as well perceived that, when using the digital ruler, they were being more precise with their measurements.

Discussion

This report presents a validation of a digital pathology system and operational workflow for remote master diagnosis, including digitization of glass slides generated from formalin fixed paraffin embedded and frozen tissue, and including hematoxylin & eosin stains, immunohistochemical stains, and special stains. All specimens were prospective clinical patient cases, first reviewed past the attending pathologists by digital means from a remote site, then the glass slides were afterward re-reviewed with a conventional brightfield microscope and verified electronically within the CLIA-licensed facility. The validation successfully demonstrated operational feasibility of supporting remote review and reporting of pathology specimens and verification of remote admission performance and usability for remote master diagnostic signout. Our findings are similar to the results of earlier studies reporting that WSI is non-inferior to glass slide review for rendering pathology diagnoses [i,2,three,4,5,6,7,eight,9,ten,11,12,13,fourteen,15,sixteen,17,xviii,19,20,21,22,23,24,25], and the electric current study represents the showtime in the United States to show acceptable concordance for master diagnosis using a digital pathology organization in a non-CLIA certified remote setting.

The COVID-19 pandemic has fueled interest in digital pathology, given the requirement for pathologists to review and report pathology for continuity of patient intendance is crucial, and protecting healthcare personal condom is key to ensuring uninterrupted, skillful pathology practice. CLIA '88 requires reporting of patient pathology to exist conducted in a CLIA-certified clinical laboratory [42 The statesC. 263a]. Hitherto, only 2 digital pathology systems have been cleared by the Nutrient and Drug Assistants for primary diagnosis in surgical pathology, although not for remote utilise. Food and Drug Administration clearance, or lack thereof, pertains to manufacturer intended apply and marketing; however, it does not govern healthcare practice. Nevertheless, CLIA '88 requirements foreclose pathologists from routinely using these systems remotely from a non-CLIA certified site. As a result, digital pathology systems used for clinical purposes take been limited to apply within CLIA-certified pathology laboratories. Due to the public wellness emergency, the Centers for Medicare & Medicaid Services issued a memorandum on March 26, 2020, applying enforcement discretion to allow pathologists to review and written report WSIs remotely, temporarily decoupling the display device from the traditionally cleared end-to-cease digital pathology organisation. The CMS memorandum states that laboratories choosing to utilize temporary testing sites (due east.k., for remote review and reporting of glass slides/WSIs) may do so if sure criteria outlined in the memorandum are met [29]. On Apr 24, 2020 the Food and Drug Assistants included further guidance pertaining to the public wellness emergency, issuing nonbinding recommendations for remote digital pathology systems. These recommendations stated that the Nutrient and Drug Administration would not object to marketing of non-510(grand) cleared systems and to their use in remote settings. The document states the Food and Drug Administration would not intend to enforce compliance for premarket notification submissions for a whole slide imaging system where the digital pathology arrangement does not create undue risk for patients or pathologists. The document further recommends validation of pathologist remote reviewing and reporting of patient pathology materials. The Higher of American Pathologists has published validation guidelines in 2013 for principal diagnosis using digital pathology systems, and there is pregnant literature attesting to the equivalency of WSI to glass slides [30]. The College of American Pathologists besides issued remote signout guidance during this public wellness emergency stating that pathology laboratories tin review and report pathology specimens from remote not-CLIA certified facilities during the temporary CLIA waiver with appropriate validation and corroboration to the CMS waiver regulations [31, 32].

This large preclinical validation report substantiates a distinct epitome emphasizing diagnostic concordance when remotely reviewing and reporting patient specimens using a digital pathology organisation, including non-510(k) cleared display devices, whole slide image viewer, and whole slide scanner. Equivalence, as defined in a previous study [three] includes all aspects of the clinical care setting pertaining to pathology. The College of American Pathologists validation guidelines for using whole slide imaging in the diagnostic setting state that each pathology laboratory planning on implementing a digital pathology system should perform their own validation studies for the respective intended clinical use. In keeping with these guidelines, our validation study encompassed surgical pathology slides at our institution, including frozen section tissue slides, special stains, and immunohistochemical studies. We also validated our complete workflow from the indicate of accessioning of cases in the laboratory information system to concluding reporting of patient pathology. This also included the whole slide paradigm scanning operational workflow, trainee involvement, remote access by pathologists and launching of whole slide images from within the laboratory information organisation, ordering of ancillary studies through the laboratory information system, and diagnostic reporting of all clinically relevant pathology parameters. A Q-probe study performed past the College of American Pathologists analyzed self-reported laboratory discrepancy frequencies by prospective secondary review of glass slides to be a median of 5.i%, for anatomic pathology [33]. In improver, every bit previously published and accustomed past the Nutrient and Drug Assistants in 2 other studies, the major discordance rate compared to a reference diagnosis was four.9% for whole slide images and 4.vi% using glass slides in one study [two]; and 3.6% for whole slide images and three.two% for glass slides in some other study [1]. In our study, the reference standard was designated as conventional light microscopy. The results from these two studies are congruent with the pre-gear up acceptance criteria of this validation report performed at our establishment. In this study there were no major diagnostic discrepancies and three minor discrepancies, with a major concordance rate of 100% (254/254) and overall concordance charge per unit (including minor discrepancies) of 98.8% (251/254), based on reads evaluated using conventional microscopy equally the reference diagnosis.

While this study demonstrates diagnostic equivalence for digital and glass slides, there still remain limitations for certain situations that will crave glass slide review with a microscope. For example, pathology cases requiring polarization of tissue (e.m., amyloid, oxalate crystals) remains as a technical limitation for diagnosis using digital pathology. Rhoads et al. reported limitations in estimation of microorganism evaluation with lower resolution scanning for whole slide images reviewed at 40 × (0.25 µm/pixel), 83× oil-immersion (0.17 µm/pixel), and 100× oil-immersion (0.xiv µm/pixel) [34]. In our validation study, 2 cases required special stains for microorganisms, ane lung biopsy with non-necrotizing granulomas where GMS and mucicarmine stains showed cryptococcus, and some other stomach biopsy, where the Helicobacter pylori immunohistochemical stain was negative.

Exterior the Us, laboratories have undergone digital transformation of their anatomic pathology services and published successful experiences. Fraggetta et al. reported successful validation and adoption of digital pathology, and likewise described the importance of minimizing human intervention to streamline operations [35]. Vodovnik et al. has reported consummate digital pathology validation of routine surgical pathology cases, autopsies, cytology, and frozen sections, including using remote signout [36,37,38]. Reporting was primarily performed past a single senior staff pathologist. Digital images were similarly accessed through the laboratory information system and displayed on either 14 in. laptops or desktop computers with dual 23 in. monitors for the remote and on-site digital reporting. Diagnostic concordance analysis was not performed for the cases reported remotely; nonetheless, a speed of 20 megabits per second was concluded to exist an acceptable network speed for remote signout. Additional literature supporting telepathology for diagnostic cyclopedia has been express to frozen sections or secondary consultations using either whole slide and static images, robotic, or alive streaming of digital images [39,40,41,42,43,44,45,46]. Withal, none of these studies from the United States include reporting from a not-CLIA certified location. Due to crisis-driven relaxation of regulatory requirements, this is an unprecedented opportunity to collect information for remote feasibility and utility to show safe and efficacious reporting of pathology using validated digital pathology systems, including from remote sites. It also opens doors towards continuous and routine apply of digital pathology for inter-institutional consultations and collaborations.

The College of American Pathologists provided further guidance such that for remote signout, laboratories tin use their ain discretion for validation protocols to demonstrate whether digital pathology systems will perform equally equivalent to conventional microscopy for a given intended utilise. In add-on, they recommend that laboratories tin deem washout periods of any elapsing as reasonable to abate recall bias [32]. The employ of a washout period in this study being less than the recommended two weeks was mandated based on the written report design to ensure reasonable turnaround times in delivering final reports. Remote digital diagnoses were not electronically verified remotely, with only a full draft diagnostic report entered into the laboratory information system. On-site drinking glass slide review was required prior to terminal release of reports, which limited the opportunity to introduce a lengthy washout period. Additionally, during the public health emergency, surgical pathology volumes were markedly decreased, suggesting at that place may exist challenges for validating clinical volumes of prospective clinical cases.

Importantly, pathologists participating equally readers while remotely reviewing and reporting of pathology specimens rated their signout experiences positively. These included pathologists from a range of years in practice and experience using digital pathology. In addition, pathologists used hardware bachelor at their respective non-CLIA certified location (e.chiliad., dwelling house), including real-world variables, such as internet access provider bandwidth. Remote reporting guidelines included ensuring all data was to be reviewed by deeply connecting within the institutional firewall and that whatsoever data viewed outside of the primary CLIA-licensed facility for signout was protected from accidental or intentional unauthorized disclosure. The pathologist experience survey showed high satisfaction with remote review and reporting using a digital pathology organisation. Comments discussed a need for enhanced whole slide prototype navigation (e.g., input devices) and ergonomic improvements with dual monitor workstations were suggested. Of the 2119 WSI, the median browse area was almost iii times the size of the industry standard past digital pathology hardware vendors (east.g., 15 × fifteen mm; 225 mm2), measured at 612 mm2. Pathologists will require effective navigation to review each slide with similar efficiency to glass slides.

No differences in diagnostic concordance or remote digital signout experience were identified by pathologists' demographics, digital pathology years of experience, network bandwidth, estimator monitor, or display resolution of the hardware used, including Windows and Macintosh personal computers. In our study, nosotros found no standard remote workstation configuration. Based on the pathologist's remote hardware, minimum specifications were recommended for remote viewing and reporting of patient pathology in our standard operating procedure (Supplementary Table ii). However, similar to findings reported in a previous written report [3], most pathologists subjectively mentioned that reviewing digital images took a longer fourth dimension compared with reviewing glass slides by conventional microscopy. This is likely due to poor input devices and navigation tools every bit well as overall feel and familiarity with digital signout, which claim boosted studies and future improvements to the technology. Overall, this study, as role of our laboratory validation, prove feasibility of remote signout for primary diagnosis and provides a milestone for our department and the digital pathology community.

With declaration of a public health emergency in the Us and subsequent enforcement discretions, pathologists may remotely review and report pathology specimens using a digital pathology system with appropriate validation. We report the start validation of a digital pathology organization from a remote non-CLIA certified facility and discover a major concordance rate of 100% comparing remote review and reporting of whole slide images to conventional microscopy. The validation included modifications to laboratory operational workflow to successfully support remote review and reporting of pathology specimens using digital pathology systems. As a issue of this validation and conditional approval from our state Department of Health, pathologists at our institution take initiated digital/remote review and reporting of patient pathology and will proceed to pursue this practice, as permissible by law, even subsequently the public health emergency has concluded. Remote access was finer utilized with sufficient operation for remote signout. Additional validation studies during these unprecedented times volition enable aggregation of data and offer insights that may influence the existing regulatory environs.

References

  1. Borowsky Ad, Glassy EF, Wallace WD, Kallichanda NS, Behling CA, et al. Digital whole slide imaging compared with light microscopy for chief diagnosis in surgical pathology: a multicenter, double-blinded, randomized study of 2045 cases. Arch Pathol Lab Med. 2020. https://doi.org/x.5858/arpa.2019-0569-OA. Online ahead of print.

  2. Mukhopadhyay Due south, Feldman MD, Abels Due east, Raheela Ashfaq, Beltaifa S, Cacciabeve NG, et al. Whole slide imaging versus microscopy for chief diagnosis in surgical pathology: a multicenter blinded randomized noninferiority report of 1992 cases (pivotal study). Am J Surg Pathol. 2018;42:39–52.

    Article  Google Scholar

  3. Hanna MG, Reuter VE, Hameed MR, Tan LK, Chiang S, Sigel C, et al. Whole slide imaging equivalency and efficiency report: experience at a large bookish center. Mod Pathol. 2019;32:916–28.

    Article  Google Scholar

  4. Al-Janabi S, Huisman A, Nap M, Clarijs R, van Diest PJ. Whole slide images every bit a platform for initial diagnostics in histopathology in a medium sized routine laboratory. J Clin Pathol. 2012;65:1107–11.

    Article  Google Scholar

  5. Bauer TW, Schoenfield L, Slaw RJ, Yerian L, Sun Z, Henricks WH. Validation of whole slide imaging for primary diagnosis in surgical pathology. Arch Pathol Lab Med. 2013;137:518–24.

    Commodity  Google Scholar

  6. Brunelli M, Beccari S, Colombari R, Gobbo S, Giobelli L, Pellegrini A, et al. iPathology cockpit diagnostic station: validation according to College of American Pathologists Pathology and Laboratory Quality Center recommendation at the Hospital Trust and University of Verona. Diagn Pathol. 2014;9:S12.

    Article  Google Scholar

  7. Cadet TP, Dilorio R, Havrilla 50, O'Neill DG. Validation of a whole slide imaging system for principal diagnosis in surgical pathology: a community hospital experience. J Pathol Inform. 2014;1:43.

    Article  Google Scholar

  8. Campbell WS, Lele SM, West WW, Lazenby AJ, Smith LM, Hinrichs SH. Concordance between whole slide imaging and light microscopy for routine surgical pathology. Hum Pathol. 2012;43:1739–44.

    Article  Google Scholar

  9. Cheng CL, Azhar R, Sng SH, Chua YQ, Hwang JSG, Chin JPF, et al. Enabling digital pathology in the diagnostic setting: navigating through the implementation journey in an academic medical centre. J Clin Pathol. 2016;69:784–92.

    Article  Google Scholar

  10. Fonyad Fifty, Krenac T, Nagy P, Zalatnai A, Csomor J, Sápi Z, et al. Validation of diagnostic accuracy using digital slides in routine histopathology. Diagn Pathol. 2012;seven:35.

    Article  Google Scholar

  11. Gilbertson JR, Ho J, Anthony 50, Jukic DM, Yagi Y, Parwani AV. Primary histologic diagnosis using automated whole slide imaging: a validation written report. BMC Clin Pathol. 2006;vi:4–19.

    Commodity  Google Scholar

  12. Goacher Due east, Randell R, Williams B, Treanor D. The diagnostic concordance of whole slide imaging and light microscopy. Arch Pathol Lab Med. 2017;141:151–61.

    Article  Google Scholar

  13. Houghton JP, Ervine AJ, Kenny SL, Kelly PJ, Napier SS, McCluggage WG, et al. Cyclopedia betwixt digital pathology and low-cal microscopy in general surgical pathology: a pilot report of 100 cases. J Clin Pathol. 2014;67:1052–5.

    Article  Google Scholar

  14. Jukic DM, Drogowski LM, Martina J, Parwani AV. Clinical examination and validation of master diagnosis in anatomic pathology using whole slide digital images. Curvation Pathol Lab Med. 2011;135:372–eight.

    PubMed  Google Scholar

  15. Snead DR, Tsang YW, Meskiri A, Kimani PK, Crossman R, Rajpoot NM, et al. Validation of digital pathology imaging for main histopathological diagnosis. Histopathology. 2016;68:1063–72.

    Article  Google Scholar

  16. Tabata K, Mori I, Sasaki T, Itoh T, Shiraishi T, Yoshimi N, et al. Whole slide imaging at primary pathological diagnosis: validation of whole slide imaging-based primary pathological diagnosis at twelve Japanese academic institutes. Pathol Int. 2017;67:547–54.

    Commodity  Google Scholar

  17. Al-Janabi Southward, Huisman A, Jonges GN, ten Kate FJW, Goldschmeding R, van Diest PJ. Whole slide images for primary diagnostics of urinary system pathology: a feasibility written report. J Ren Inj Prev. 2014;three:91–6.

    PubMed  Google Scholar

  18. Al-Janabi S, Huisman A, Nikkels PG, 10 Kate FJ, van Diest PJ. Whole slide images for primary diagnostics of paediatric pathology specimens: a feasibility study. J Clin Pathol. 2013;66:218–23.

    Article  Google Scholar

  19. Al-Janabi S, Huisman A, Vink A, Leguit RJ, A Offerhaus GJ, Ten Kate FJ, et al. Whole slide images for main diagnostics in dermatopathology: a feasibility report. J Clin Pathol. 2012;65:152–eight.

    Commodity  Google Scholar

  20. Al-Janabi S, Huisman A, Vink A, Leguit RJ, A Offerhaus GJ, Ten Kate FJ, et al. Whole slide images for principal diagnostics of gastrointestinal tract pathology: a feasibility study. Hum Pathol. 2012;43:702–7.

    Article  Google Scholar

  21. Al-Janabi S, Huisman A, Willems SM, Van Diest PJ. Digital slide images for main diagnostics in breast pathology: a feasibility written report. Hum Pathol. 2012;43:2318–25.

    CAS  Commodity  Google Scholar

  22. Ordi J, Castillo P, Saco A, Del Pine Thousand, Ordi O, Rodríguez-Carunchio L, et al. Validation of whole slide imaging in the chief diagnosis of gynaecological pathology in a university hospital. J Clin Pathol. 2015;68:33–nine.

    Commodity  Google Scholar

  23. Reyes C, Ikpatt OF, Nadji M, Cote RJ. Intra-observer reproducibility of whole slide imaging for the primary diagnosis of breast needle biopsies. J Pathol Inform. 2014;v:5.

    Article  Google Scholar

  24. Stathonikos Northward, Nguyen TQ, Spoto CP, Verdaasdonk MAM, van Diest PJ. Existence fully digital: perspective of a Dutch academic pathology laboratory. Histopathology. 2019;75:621–35.

    Commodity  Google Scholar

  25. Evans AJ, Salama ME, Henricks WH, Pantanowitz L. Implementation of whole slide imaging for clinical purposes: problems to consider from the perspective of early adopters. Arch Pathol Lab Med. 2017;141:944–59.

    Article  Google Scholar

  26. Clinical Laboratory Improvement Amendments of 1988 (CLIA) Title 42: The Public Health and Welfare. Subpart ii: Clinical Laboratories (42 U.Southward.C. 263a). https://world wide web.govinfo.gov/content/pkg/USCODE-2011-title42/pdf/USCODE-2011-title42-chap6A-subchapII-partF-subpart2-sec263a.pdf. Accessed 22 Apr. 2020.

  27. U.Due south. Food and Drug Administration. FDA news release: FDA allows marketing of first whole slide imaging system for digital pathology. https://www.fda.gov/drugs/informationondrugs/approveddrugs/ucm553358.htm. Accessed 28 Mar 2020.

  28. U.Due south. Nutrient and Drug Administration. Leica Biosystems Receives FDA Clearance for Aperio AT2 DX Digital Pathology Organization https://www.accessdata.fda.gov/cdrh_docs/pdf19/K190332.pdf. Accessed 28 Mar 2020.

  29. Centers for Medicare & Medicaid Services Memorandum, Clinical Laboratory Improvement Amendments (CLIA) Laboratory Guidance During COVID-nineteen Public Wellness Emergency, March 26, 2020, https://www.cms.gov/medicareprovider-enrollment-and-certificationsurveycertificationgeninfopolicy-and-memos-states-and/clinical-laboratory-improvement-amendments-clia-laboratory-guidance-during-covid-19-public-wellness. Accessed 28 Mar 2020.

  30. Pantanowitz L, Sinard JH, Henricks WH, Fatheree LA, Carter AB, Contis L, et al. Validating whole slide imaging for diagnostic purposes in pathology: guideline from the Higher of American Pathologists Pathology and Laboratory Quality Center. Arch Pathol Lab Med. 2013;137:1710–22.

    Article  Google Scholar

  31. College of American Pathologists. COVID-19—Remote Sign-Out Guidance. 2020. https://documents.cap.org/documents/COVID19-Remote-Sign-Out-Guidance-vFNL.pdf. Accessed 30 Mar 2020.

  32. College of American Pathologists. Remote Sign-Out FAQs. 2020. https://documents.cap.org/documents/Remote-Sign-Out-FAQs-Last.pdf Accessed 3 Apr. 2020.

  33. Raab SS, Nakhleh RE, Red SG. Patient safe in anatomic pathology: measuring discrepancy frequencies and causes. Curvation Pathol Lab Med. 2005;129:459–66.

    PubMed  Google Scholar

  34. Rhoads DD, Habib-Bein NF, Hariri RS, et al. Comparing of the diagnostic utility of digital pathology systems for telemicrobiology. J Pathol Inform. 2016;7:10.

    Article  Google Scholar

  35. Fraggetta F, Garozzo S, Zannoni GF, Pantanowitz L, Rossi ED. Routine digital pathology workflow: the Catania experience. J Pathol Inform. 2017;8:51.

    Article  Google Scholar

  36. Vodovnik A. Distance reporting in digital pathology: a study on 950 cases. J Pathol Inform. 2015;6:18.

    Article  Google Scholar

  37. Vodovnik A, Aghdam MR, Espedal DG. Remote autopsy services: a feasibility study on nine cases. J Telemed Telecare. 2018;24:460–4.

    Article  Google Scholar

  38. Vodovnik A, Aghdam MRF. Complete routine remote digital pathology services. J Pathol Inform. 2018;9:36.

    Article  Google Scholar

  39. Pantanowitz Fifty, Wiley CA, Demetris A, Lesniak A, Ahmed I, Cable West, et al. Feel with multimodality telepathology at the University of Pittsburgh Medical Centre. J Pathol Inform. 2012;3:45.

    Article  Google Scholar

  40. Thrall MJ, Rivera AL, Takei H, Powell SZ. Validation of a novel robotic telepathology platform for neuropathology intraoperative touch preparations. J Pathol Inform. 2014;five:21.

    Commodity  Google Scholar

  41. Wilbur DC, Madi Grand, Colvin RB, Duncan LM, Faquin WC, Ferry JA, et al. Whole-slide imaging digital pathology equally a platform for teleconsultation: a airplane pilot report using paired subspecialist correlations. Arch Pathol Lab Med. 2009;133:1949–53.

    PubMed  PubMed Cardinal  Google Scholar

  42. Sirintrapun SJ, Rudomina D, Mazzella A, Feratovic R, Alago West, Siegelbaum R, et al. Robotic telecytology for remote cytologic evaluation without an on-site cytotechnologist or cytopathologist: a tale of implementation and review of constraints. J Pathol Inform. 2017;8:32.

    Article  Google Scholar

  43. Chong T, Palma-Diaz MF, Fisher C, Gui D, Ostrzega NL, Sempa G, et al. The California Telepathology Service: UCLA'southward experience in deploying a regional digital pathology subspecialty consultation network. J Pathol Inform. 2019;10:31.

    Article  Google Scholar

  44. Chen J, Jiao Y, Lu C, Zhou J, Zhang Z, Zhou C. A nationwide telepathology consultation and quality control program in China: implementation and result analysis. Diagn Pathol. 2014;9(Suppl 1):S2.

    Article  Google Scholar

  45. Têtu B, Perron É, Louahlia South, Paré G, Trudel MC, Meyer J. The Eastern Québec Telepathology Network: a iii-yr experience of clinical diagnostic services. Diagn Pathol. 2014;9(Suppl ane):S1.

    Commodity  Google Scholar

  46. Mpunga T, Hedt-Gauthier BL, Tapela North, Nshimiyimana I, Muvugabigwi G, Pritchett N, et al. Implementation and validation of telepathology triage at cancer referral center in Rural Rwanda. J Glob Oncol. 2016;2:76–82.

    Commodity  Google Scholar

Download references

Funding

This enquiry was funded in function through the NIH/NCI Cancer Center Support Grant P30 CA008748.

Author information

Affiliations

Contributions

MGH, VER, OA, DK, DSK, MH: concept and design, information collection, data analysis/estimation, manuscript drafting/revision. PS, JP, ES, MAF, PN, ML, CE: concept and design, data analysis/interpretation, manuscript revision. SJS, KB, JLS, EB, LKT, BX, TB, NA, LHH, LHE: data collection, manuscript revision.

Respective writer

Correspondence to Matthew G. Hanna.

Ethics declarations

Conflict of involvement

MGH is a consultant for Paige.AI and an counselor to PathPresenter. PJS is a co-founder and equity holder of Paige.AI. DSK is a founder, consultant, and equity holder for Paige.AI. All other authors declare no conflicts of interest.

Additional information

Publisher'south annotation Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary data

Rights and permissions

About this article

Verify currency and authenticity via CrossMark

Cite this article

Hanna, K.G., Reuter, V.E., Ardon, O. et al. Validation of a digital pathology system including remote review during the COVID-nineteen pandemic. Mod Pathol 33, 2115–2127 (2020). https://doi.org/10.1038/s41379-020-0601-v

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Effect Date:

  • DOI : https://doi.org/10.1038/s41379-020-0601-5

Further reading

merrittsuaid1951.blogspot.com

Source: https://www.nature.com/articles/s41379-020-0601-5

0 Response to "Pathology Service Reading Out Biopsy Slides in Academia"

Enviar um comentário

Iklan Atas Artikel

Iklan Tengah Artikel 1

Iklan Tengah Artikel 2

Iklan Bawah Artikel